Agilent Technologies N5700 Manuel D’Utilisation

Page de 128
SCPI Programming Commands – At a Glance. 
Series N5700 User’s Guide 
15 
SCPI Programming Commands – At a Glance
.
 
NOTE 
Some [optional] commands have been included for clarity. Chapter 6 contains a 
complete description of all programming commands. 
Subsystem Commands. 
SCPI Command 
Description 
ABORt 
Aborts the triggered action 
 
 
CALibrate  
    :CURRent[:LEVel] 
Calibrates the output current programming 
    :DATA <NRf> 
Enters the calibration value 
    :DATE “<date>” 
Sets the calibration date 
    :LEVel P1 | P2 
Advances to the next calibration step 
    :PASSword <NRf> 
Sets the numeric calibration password 
    :STATE <Bool> [,<NRf>] 
Enables/disables calibration mode 
    :VOLTage[:LEVel] 
Calibrates the output voltage programming 
 
 
INITiate  
    [:IMMediate][:TRANsient] 
Initiates the trigger system 
    :CONTinuous[:TRANsient] 
Enables/disables continuous triggers  
 
 
MEASure  
    [:SCALar] 
 
          :CURRent[:DC]?  
Returns the measured output current 
          :VOLTage[:DC]?  
Returns the measured output voltage 
 
 
OUTPut  
    [:STATe] <Bool> 
Enables/disables the specified output  
    :PON 
 
          :STATe RST | AUTO 
Programs the Power-On State 
    :PROTection 
 
          :CLEar   
Resets latched protection 
 
 
[SOURce:]  
    CURRent 
 
          [:LEVel] 
 
                 [:IMMediate][:AMPLitude] <NRf+>  
Sets the output current 
                 :TRIGgered[:AMPLitude] <NRf+>  
Sets the triggered output current 
          :PROTection  
 
                :STATe <Bool>  
Enables/disables over-current protection  
    VOLTage 
 
          [:LEVel] 
 
                 [:IMMediate][:AMPLitude] <NRf+> 
Sets the output voltage 
                 :TRIGgered[:AMPLitude] <NRf+> 
Sets the triggered output voltage 
          :LIMit  
 
                 :LOW <NRf+> 
Sets the low-voltage limit 
          :PROTection 
 
                 [:LEVel] <NRf+> 
Sets the over-voltage protection level