Emerson P/N MMI-20019808 Manuel D’Utilisation
40
Micro Motion EtherNet/IP Module
Device Profile
B.3.2
Output assembly for standard configuration
B.3.3
Input parameters (explicit data) for standard configuration
To update any of these attribute values, the associated trigger byte must be toggled before
reading the attribute value (executing the Get service). See Section B.3.4 for more information
on trigger bytes.
reading the attribute value (executing the Get service). See Section B.3.4 for more information
on trigger bytes.
36–39
Get
Gas standard volume total
Float
Valid only when Gas Standard Volume is
enabled.
enabled.
Byte
Access
Name
Type
Notes
0–3
Get/Set
External Temperature
Float
4–7
Get/Set
External Pressure
Float
8
Get/Set
Start/Stop Totals
Byte
0: Stop
1: Start
1: Start
9
Get/Set
Reset All Process Totals
Byte
0: No action
1: Reset
1: Reset
10
Get/Set
Reset All Inventory Totals
Byte
0: No action
1: Reset
1: Reset
11
Get/Set
Start Zero
Byte
0: Abort or no action
1: Start
1: Start
12–13
Get/Set
Start Smart Meter Verifica-
tion
tion
Word
See
Class
Instance Attribute Access Name
Type
Trigger
Byte Write
Attribute
Byte Write
Attribute
Description
B0h
01h
01h
Get
Mass flow rate unit
U16
0Ch
See
02h
Get
Density unit
U16
0Ch
See
03h
Get
Temperature unit
U16
0Ch
See
04h
Get
Volume flow rate unit
U16
0Ch
See
05h
Get
Pressure unit
U16
0Ch
See
06h
Get
Mass total/inventory unit
U16
0Ch
See
07h
Get
Volume total/inventory unit
U16
0Ch
See
08h
Get
Zero time
U16
0Dh
Seconds
09h
Get
Standard deviation of auto zero Float
0Eh
0Ah
Get
Present flow signal offset at
zero flow
zero flow
Float
0Eh
0Bh
Get
Failed Zero Calibration Value
Float
0Eh