Emerson P/N MMI-20019808 Manuel D’Utilisation

Page de 78
 
40
Micro Motion EtherNet/IP Module
Device Profile
B.3.2
Output assembly for standard configuration
B.3.3
Input parameters (explicit data) for standard configuration
To update any of these attribute values, the associated trigger byte must be toggled before 
reading the attribute value (executing the Get service). See Section B.3.4 for more information 
on trigger bytes.  
36–39
Get
Gas standard volume total
Float
Valid only when Gas Standard Volume is 
enabled.
Byte
Access
Name
Type
Notes
0–3
Get/Set
External Temperature
Float
4–7
Get/Set
External Pressure
Float
8
Get/Set
Start/Stop Totals
Byte
0: Stop
1: Start
9
Get/Set
Reset All Process Totals
Byte
0: No action
1: Reset
10
Get/Set
Reset All Inventory Totals
Byte
0: No action
1: Reset
11
Get/Set
Start Zero
Byte
0: Abort or no action
1: Start
12–13
Get/Set
Start Smart Meter Verifica-
tion
Word
See 
Class
Instance Attribute Access Name
Type
Trigger 
Byte Write
Attribute
Description
B0h
01h
01h
Get
Mass flow rate unit
U16
0Ch
Se
02h
Get
Density unit
U16
0Ch
Se
03h
Get
Temperature unit
U16
0Ch
Se
04h
Get
Volume flow rate unit
U16
0Ch
Se
05h
Get
Pressure unit
U16
0Ch
Se
06h
Get
Mass total/inventory unit 
U16
0Ch
Se
07h
Get
Volume total/inventory unit 
U16
0Ch
Se
08h
Get
Zero time
U16
0Dh
Seconds
09h
Get
Standard deviation of auto zero Float
0Eh
0Ah
Get
Present flow signal offset at 
zero flow
Float
0Eh
0Bh
Get
Failed Zero Calibration Value
Float
0Eh