Emerson P/N MMI-20019808 Manuel D’Utilisation

Page de 78
Device Profile
User Manual
 
43
B.3.5
Input assembly for concentration measurement configura-
tion
Byte
Access
Name
Type
Notes
0–3
Get
Mass flow rate
Float
4–7
Get
Density
Float
8–11
Get
Temperature
Float
12–15
Get
Volume flow rate (liquid 
volume)
Float
16–19
Get
Mass total
Float
20–23
Get
Standard volume flow
Float
24–27
Get
Standard volume total
Float
28–31
Get
Standard volume inventory
Float
32–35
Get
Net mass flow
Float
36–39
Get
Net mass total
Float
40–43
Get
Netmass inventory
Float
44–47
Get
Net volume flow
Float
48–51
Get
Net volume total
Float
52–55
Get
Net volume inventory
Float
56–59
Get
Reference density
Float
60–63
Get
Specific gravity
Float
64–67
Get
Concentration
Float
68–71
Get
Density (fixed Baume units)
Float
72–75
Get
Volume total (liquid)
Float
76–79
Get
Drive gain
Float
80–81
Get
Status word
U16 or Word
• For Model 1700 Analog, Model 2700 
Analog, Model 1500 Analog, Model 
2500, and all Series 3000 transmitters: 
SNS Status Word 1 (see Section B.7.1)
• For MVD Direct Connect and 9739 MVD 
transmitters: SNS Status Word 2 (see 
Section B.7.2)
82–83
Get
Status word
U16 or Word
• For Model 1700 Analog, Model 2700 
Analog, Model 1500 Analog, Model 
2500, and all Series 3000 transmitters: 
SNS Status Word 2 (see Section B.7.2)
• For MVD Direct Connect and 9739 MVD 
transmitters: SNS Status Word 1 (see 
Section B.7.1)