HP (Hewlett-Packard) RM500SL Manuel D’Utilisation

Page de 112
0611 
RM500SL User’s Guide Version 2.8  
Page 19 
3.6 
Screen messages 
FastFacts 3.6: Screen messages 
3.7 
Program memory removal and replacement  
FastFacts 3.7: Program memory removal and replacement 
The Title bar (top line of the display screen) informs you of the selected Test (e.g., 
Multicurve). 
The Message bar (bottom line of the display screen) suggests the next step in a 
test or informs you of the state of the instrument or conditions that could affect your 
data. For example, it will inform you if the microphones need to be calibrated. 
Operating software is contained on a USB flash drive located in the test chamber 
under the printer paper roll cradle as shown. Software updates are shipped on a 
CD-ROM and must be transferred to the USB flash drive in order to be used. This 
requires that the USB flash drive be removed from the RM500SL and plugged into 
the USB port on a PC that has a CD-ROM drive. 
To remove the USB flash drive, ensure that the RM500SL power is off and: 
1)  Open the test chamber and remove the foam from the test chamber floor. 
2)  Pull the USB flash drive straight toward the front of the unit until it is free. 
To install the USB flash drive, ensure that the RM500SL power is off and: 
1)  Open the test chamber and remove the foam from the test chamber floor. 
2)  Gently insert the USB flash drive into the USB port, with its foam pad down. 
3)  Replace the foam in the floor of the test chamber. 
NOTE that the USB 
flash drive has been 
especially configured 
for the RM500SL. 
Replacing this flash 
drive with another 
can cause erratic 
operation. Such 
replacement should 
be done only when 
authorized and 
ONLY by an 
Audioscan service 
representative.