HP (Hewlett-Packard) RM500SL Manuel D’Utilisation

Page de 112
0611 
RM500SL User’s Guide Version 2.8  
Page 34 
7.9 
Calibration of HIT reference microphone 
FastFacts 7.9: Calibration of HIT reference microphone 
Remove any coupler from the coupler microphone before calibration. 
1)  Position the HIT reference microphone opening concentric with the coupler 
microphone opening and about 1 – 2 mm from as shown below. 
2)  Close and seal the test chamber by tightening the latch at the front of the lid. 
3)  Press <Tests>, then highlight and <PICK> [Calibrate] from the Hearing 
Instruments Test column. 
4)  Highlight, then <PICK> [Daily] or [Weekly] to select the calibration interval. 
5)  Highlight and <PICK> [Calibrate]. 
See 7.8: HIT calibration facts
 
It is recommended that you print a copy of the initial calibration curve and compare 
it with your daily or weekly curve. Any change in this curve indicates a drift in 
either the coupler or the reference microphone which needs further investigation. 
See 
7.10: Calibration check for coupler microphone
 for an additional check.