8.5 
ANSI 2003 Linear and AGC tests 
FastFacts 8.5: ANSI 2003 Linear and AGC tests 
1)  Press <Tests>, then highlight and <PICK> [Linear] or [AGC]. 
2)  Highlight and <PICK> [Start test]; test setup instructions will be displayed. 
3)  Refer to 
7.11: Coupling the hearing instrument
7.12: Positioning the hearing 
instrument
 and 
2.5: Battery pill use and care
 to prepare the instrument for 
testing. 
4)  Turn the instrument ON
5)  Set the gain control (manual or programmable) to maximum with all other 
controls set to provide the widest frequency response, greatest gain and 
highest maximum output. AGC controls and/or programmable parameters 
should be set to minimize AGC action (AGC off or threshold high, compression 
ratio close to 1). 
6)  Press <Continue> to run the test or press <Cancel> to abort the test. Note that, 
if required, the RM500SL will pause during the test to request that the gain 
control be adjusted to the Reference Test Setting (RTS). If necessary, open the 
HIT chamber and, without moving the hearing instrument, adjust the gain 
control until the large vertical arrow on the screen prompt disappears. Press 
<Continue> to resume the test. 
During the AGC test, the RM500SL will pause to request that AGC controls or 
programmable parameters be set for maximum AGC effect. If the manufacturer 
does not provide settings to produce maximum AGC effect, use settings that result 
in the highest gain for soft sounds and the lowest gain for loud sounds. This will 
typically result in the lowest compression threshold, a flat I/O curve and the 
longest attack/release times. These are not “typical” characteristics but indicate 
the adjustment range possible for these parameters. 
If you are unsure which test sequence to run, try [AGC] first. If attack and release 
times are all near zero, the instrument is probably linear.