16.6  REAR measurement procedure 
FastFacts 16.6: REAR measurement procedure 
1)  Press <Tests>, then highlight and <PICK> [Insertion gain] from the REM list 
2)  Referring to 
10: Real-Ear Measurement Setup
 and 
11: REM Screen Setup
prepare the system, position the patient and insert the probe tube. (For tight-
fitting hearing aids or earmolds, the probe tube may be inserted through a vent.) 
3)  Without disturbing the probe tube, insert the hearing instrument or earmold. 
4)  Referring to 
16.4: Audiometric data entry
, enter threshold data. 
5)  Highlight and <PICK> button 1, 2, or 3 in the REAR/REIG column. This will 
present a signal from the REM speaker and display the REAR Setup box. 
6)  Highlight and <PICK> [Stimulus], then highlight and <PICK> [Pink noise] or 
[Swept]. Use pink noise when using the NAL-NL1 fitting formula. Use Swept 
when testing maximum output capabilities. 
7)  Highlight and <PICK> [Level], then highlight and <PICK> a stimulus level from 
the list box. Use the level shown on the target REIG curves. If no level is 
shown, use the lowest level required to overcome background noise. 
8)  Adjust the hearing instrument so that the REIG curve (SPL scale) or Aided 
threshold curve (HL scale) approximates the dotted target curve. For multi-level 
targets, match each curve using the stimulus level indicated. 
9)  Press <Continue> to complete the test and capture the data. 
See 
16.2: Insertion gain in SPL
 and 
16.3: Insertion gain in HL
 for an explanation of 
screen curves. 
See 16.7:SII calculation in Insertion gain 
NOTE: Because it is analyzed in 1/3 octave bands, the Noise stimulus produces a 
REAR up to 12 dB below that produced by the Swept stimulus at the same overall 
level. The REIGs will be similar except at levels which cause non-linear operation 
of the hearing instrument.