HP (Hewlett-Packard) RM500SL Manuel D’Utilisation

Page de 112
0611 
RM500SL User’s Guide Version 2.8  
Page 75 
17.9  REM or S-REM mode 
FastFacts 17.9: REM or S-REM mode 
The Speechmap environment provides two hearing instrument verification options: 
REM (real-ear measurement) and S-REM (simulated REM). In REM mode, the 
stimulus is controlled by the REM reference microphone below the ear and SPL in 
the ear canal is measured by the probe microphone. In S-REM mode, levels 
measured in the 2cc coupler are converted to estimated ear canal levels using the 
RECD values (age-appropriate average, entered, or measured). In addition, the 
stimulus delivered to the hearing instrument microphone is modified to include 
average microphone location effects for the hearing instrument type indicated. The 
screen display for both modes is the same, showing SPL, measured or simulated, 
in the ear canal. 
For REM mode, review 
10: Real-Ear Measurement Setup
For S-REM mode review 
7.9: Calibration of HIT reference microphone
7.11: 
Coupling the hearing instrument, 7.12: Positioning the hearing instrument. 
It is suggested that hearing instruments be preset in S-REM mode with fine tuning 
done in REM mode. 
v  Highlight and <PICK> [Mode] to toggle between REM and S-REM. 
Note that REM mode should be used for vented or open fittings. Vent effects 
depend on the vent, the ear and the hearing aid settings.