HP (Hewlett-Packard) RM500SL Manuel D’Utilisation

Page de 112
0611 
RM500SL User’s Guide Version 2.8  
Page 81 
18.6  Fitting to targets for loud sounds 
FastFacts 18.6: Fitting to targets for loud sounds 
4)  Select REAR 2 & set stimulus to [MPO]. In S-REM mode, a 90 dB tone sweeps 
continuously. In REM mode, the tone sweeps when you press <Continue>
5)   In S-REM mode, adjust high-level gain and/or output limiting (e.g.,MPO, PC) 
so the curve matches the + REAR90 targets (DSL) or approaches but doesn’t 
exceed the 
∗ RESR targets (NAL-NL1) at all frequencies. Press <Continue> to 
record the MPO curve. In REM mode, advise the client to point to the screen if 
a sound becomes too loud during the test (see 
10.5: Max TM SPL setup)
. Press 
<Continue> to run and record a single sweep. Adjust the gain and output for 
loud sounds as for S-REM, factoring in client response. Repeat as necessary.