20.8  Invalid HIT calibration 
FastFacts 20.8: Invalid HIT calibration 
20.9  No REM ref. mic. detected  
FastFacts 20.9: No REM ref. mic. detected 
20.10 Invalid REM calibration 
FastFacts 20.10: Invalid REM calibration 
No REM reference mic. detected. 
Check connection on the REM probe microphone. 
v  Check that a probe microphone assembly is securely plugged into the PROBE 
socket to the left of the test chamber. 
v  Unplug the HIT reference microphone in the test chamber and plug the REM 
probe microphone in its place. Try HIT calibration using the reference 
microphone of the REM probe microphone in place of the HIT reference 
microphone. If this is fails, the REM probe microphone is defective and must 
be replaced. 
Invalid REM Calibration! 
This message may be displayed during the REM calibration procedure. It indicates 
that the calibration curve is significantly different from what is expected. It may be 
accompanied by a REM reference microphone error. 
v  Check that a probe microphone assembly is securely plugged into the PROBE 
socket to the left of the test chamber. 
v  Try using a new probe tube. Ensure it is correctly attached and positioned. 
v  Try the steps in 
20.9: No REM ref. mic. detected
v  Unplug the coupler microphone from the test chamber and plug the REM 
probe microphone in its place. Position the open end of the probe tube near 
the HIT reference microphone in the test chamber (use putty to hold it) and 
close the test chamber lid. Run a Multicurve test at a stimulus level of 70 dB 
SPL. The result should have the shape of a normal REM calibration curve. If it 
does not, the REM microphone is defective and must be replaced. 
Invalid HIT Calibration! 
This message may be displayed during the HIT calibration procedure. It indicates 
that the calibration curve is significantly different from what is expected. It may be 
accompanied by a missing HIT reference microphone error. 
v  Check that the there is no coupler on the coupler microphone and that the 
microphones are correctly positioned for HIT calibration. 
v  Check that the coupler microphone is securely plugged into the correct socket. 
v  Follow the suggestions in 
20.7: No HIT reference mic. detected