14
Instruction Manual V1.2 - 11/2006
2.
Safety
2.3.9 Infectious/radioactive material
Danger!
Use caution when working with potentially
infectious specimens:
-->Risk of infection!
When working with potentially infectious /
radioactive specimens:
Wear protective clothes (gloves, protec-
tive boots, mask, lab coat), in compliance
with radiation safety regulations and/or
in-house regulations on handling infec-
tious / radioactive material.
When working with radioactive specimens:
Comply with applicable radiation safety
regulations!
Dispose of radioactive specimen waste ac-
cording to applicable regulations.
2.3.10 Disinfection and cleaning
Prior to disinfection, switch the instrument
off and unplug it from mains.
For removal of the microtome from the cryo-
chamber see chapter 2.3.11 ‘ Removing the
microtome’.
For disinfection, wear protective gear:
(gloves, mask, lab coat etc.)!
For disinfection, only use alcohol-based dis-
infectants!
Do not use solvents (xylene, acetone etc.)
for cleaning or disinfection!
Do not spray disinfectants into the evapora-
tor!
--> Risk of icing!
Explosion hazard when working with alcohol:
Make sure the premises are appropriately
ventilated!
When using disinfectants and detergents,
comply with all safety instructions supplied
by the manufacturer of the product!
Dispose of waste liquids from disinfection/
cleaning as well as of sectioning waste ac-
cording to applicable regulations on disposal
of special category waste!
Disinfected accessories must be thoroughly
dry when reinserting them into the chamber!
--> Risk of icing!
Make sure the chamber is completely dry
before switching the instrument back on:
--> Explosion hazard through alcohol vapors!
--> Risk of icing!