Leica CM3050 S Manuel D’Utilisation

Page de 66
31
Leica CM3050 S – Cryostat
The automatic defrost cycle takes place once
every 24 hours at the selected defrost time.
7. Duration of defrost cycle:
Press key until menu item ‘Defrost period’
is displayed:
Use defrost keys (
 or 
) to adjust de-
frost time. (Only if necessary - see below).
Important information for setting the
defrost parameters
Starting time of defrost cycle:
We recommend to program the
automatic defrost cycle for a time out-
side regular working hours as a
period of 4 hours should pass after the
defrost cycle before work with the
instrument is resumed (see also
chapter 5.7.1 ‘Automatic chamber
defrost cycle’).
Defrost cycle duration:
The duration of the chamber defrost
cycle is set to 8 minutes in factory.
This setting should only be
changed, if the evaporator is not
completely de-iced during the
8-minute cycle.
In such cases, increase the defrost
cycle duration gradually until the the
evaporator is ice-free after defrosting.
The selected defrost cycle duration is
valid for both automatic and
manual chamber defrost cycle.
4. Quick-freezing at maximum specimen
temperature* (instruments with specimen
cooling only)
Press key until menu item ‘Fast object
freezing’ is displayed:
or
Press lock/unlock key to activate or de-
activate (ON/OFF).
5. Real time
Press key until menu item ‘Time’ is dis-
played:
Use arrow keys (
 or 
) to set real time.
Setting minutes as well as hours in
1-minute steps. Holding arrow button in-
creases scrolling speed!
6. Starting time of automatic defrost cycle:
Press key until menu item ‘Defrost time’ is
displayed:
(Time on display = start time of defrost
cycle)
Use arrow keys (
 or 
) to select the de-
sired defrost start time.
5.
Operation
*) with 90° prism - see also chapter 5.5.2.
F A S T
O B J
F R E E Z I N
O F F
F A S T
O B J
F R E E Z I N
O N
T I M E
1 0 : 2 3
D E F R
T I M E
4 : 3 0
D E F R O S T
P E R I O D
C H A M B E R :
0 8
M I N