Leica CM3050 S Manuel D’Utilisation

Page de 66
48
 Instruction Manual V1.1 02/2000
6.
Troubleshooting, applications tips
6.2
Potential problems - causes and remedies
Remedies
- Eliminate draft sources or
change place of installation of
the cryostat.
- Close sliding window.
- Consider wearing a mask.
- Select lower temperature.
- Wait until knife and/or anti-roll
plate have reached chamber
temperature.
- Select higher temperature.
- Eliminate cause.
- Select lower temperature.
- Trim specimen parallel; in-
crease section thickness.
- Readjust anti-roll plate.
- Align correctly.
- Set correct clearance angle.
- Use different part of knife or
replace knife.
- Clean with dry cloth or brush.
- Replace anti-roll plate.
- Use different part of knife or
replace knife.
Causes
- Cryostat is exposed to draft
(open windows, doors, air-
conditioning).
- Sliding window has been open
for a long period of time at a
very low chamber temperature.
- Frost build-up caused by
breathing into the cryocham-
ber.
- Specimen not cold enough.
- Knife and/or anti-roll plate not
yet cold enough - sections
melt.
- Specimen too cold.
- Static electricity / draft.
- Specimen not cold enough.
- Large surface specimen.
- Anti-roll plate poorly adjusted.
- Anti-roll plate poorly aligned
relative to knife edge.
- Wrong clearance angle
selected.
- Knife blunt or damaged.
- Knife and/or anti-roll plate dirty.
- Edge of anti-roll plate dam-
aged.
- Blunt knife.
Problem
Frost on chamber walls and
microtome
Sections smear
Sections splinter
Sections not properly flattened
Sections not properly flattened
despite correct temperature
and correctly aligned anti-roll
plate