Leica CM 3050 S – Cryostat
Removal
The lamp (3) is located behind the glare
shield (4).
Carefully hold lamp with both of your hands.
Slightly pull the lamp downward at the left
side to remove it from fastening clip (5).
Pull the lamp towards the left to remove it
from its socket (6).
Installing the replacement lamp
Hold the lamp in correct position (see draw-
ing below) and push towards the right until it
locks into the socket.
Push the lamp slightly upwards until it locks
into the clip.
Put the sliding window back in place.
Reconnect the instrument to mains and
switch it back on.
7.6
Exchanging the fluorescent light lamp
Prior to exchanging the lamp,
please read and observe the fol-
lowing safety instructions:
• ‘Maintenance’
(Chapter 2.3.13)
Removing the sliding window
Hold sliding window (1) at grip bar (2), lift
slightly and pull out towards yourself.
7.
Disinfection, cleaning and maintenance