Leica DM4000B Manuel D’Utilisation

Page de 69
27
5. Assembly
ICT/P transmitted light polarizer
• Using the left clamping screw, fasten the ICT/P
transmitted light polarizer to the underside of
the condenser holder (Fig. 27).
• Make sure that the red index point on the front
of the polarizer is aligned with 0.
• If necessary, insert the compensators (
λ
- and
λ
/4 plates) into the polarizer’s receptacle
(Fig. 28).
Incident light polarizers:
R/P polarizer, rotating polarizer
L/ICR, R/ICR polarizer
• Remove the plug cap on the right side of the
incident light axis (Fig. 29).
• Insert the polarizer into the receptacle until it
latches in place.
Motorized polarizer
• A motorized polarizer is already installed and
ready for operation in the DIC condenser.
Transmitted light and incident light analyzer
Fig. 27
Assembly of the ICT/P transmitted light polarizer
1
Clamping screw
Fig. 28
Inserting the compensators
Fig. 29
Inserting the polarizer
1
The plug cap is replaced with the polarizer.
1
1