6. Assembly
!
Fig. 3
Specimen stage with specimen holder
1
Lock screws for specimen holder
The microscope components are logically as-
sembled in this order:
• Stage
• Condenser
• Fluorescence illuminator*
• Intermediate systems*
• Tube
• Eyepieces
• Objectives
• Lamp housings with light sources*
• Polarization equipment*
Only one commonly used screwdriver is
necessary for assembly, which is included in the
delivery package.
The tool can be stored on a magnetic retainer
on the underside of the stage at the right.
When using intermediate systems and optical
accessories, the sequence may vary.
In this case, read Chapter
"6.9 Optional Accessories" →
 
p. 25.
6.1 Stage
Caution:
Before completing the stage, make sure no ob-
jectives are installed!
Remove the srew located under the stage in the
front.
Specimen Holder
• Place the specimen holder on the stage and
fasten it with the two screws (3.1)
Coaxial Pinion
Note:
The coaxial pinion can be mounted on the left-
or right-hand side.
• First, place the flat fine focus wheel on the
side to which you intend to mount the coaxial
pinion; the wheel is held in place magnetically
(4.1);  ensure that the button snaps into place.
Attach the other focus knob on the opposite side
• Loosen the lock screw (5.1) at the front left-
hand side of the stage
• Slide the stage as far back as possible
• Attach the coaxial pinion with the srew (6.1)
• Return the stage to the starting position and
retighten the lock screw; after installation of
the stage control, move object guide all the
way to the left side of the instrument; keep
turning when guide has reached the end of
travel until a click noise is heard
1
6. Assembling the Microscope