Leica VT1200S Manuel D’Utilisation

Page de 58
8
Operating Manual V1.1 – 05/2007
3.
Instrument Characteristics
3.1
Technical Data for the VT1200
General data:
Cutting frequency (± 10 %) ................................................................................................................. 85 Hz (± 10%)
Amplitude .................................................................................................... from 0 – 3 mm, in 0.05 mm increments
Cutting speed (± 10 %) ...................................................................................................................... 0.01 – 1.5 mm/s
Return speed (± 10 %) ................................................................................................................................. 2.5 mm/s
Total vertical specimen stroke ................................................................................................. 20 mm (motorized)
Cutting range ............................................................................................................................... 45 mm (adjustable)
Maximum specimen size:
With standard blade holder ............................................................................................................. 33 x 50 mm
Specimen orientation, rotating ................................................................................................................... 360 °
Specimen plate, swiveling ...................................................................................................................... 0 – 10 °
Section thickness adjustment .................................................................................. manual, in 1 μm increments
 Ambient conditions:
Operating temperature range: .......................................................................................... min. 10 °C – max. 35 °C
Relative humidity: ....................................................................................................................................... max. 60 %
Storage temperature: ................................................................................................................................... 5 – 55 °C
Storage humidity: ............................................................................................................................................. < 60 %
Height: ........................................................................................................................ up to 2000 m above sea level
Electrical data:
Rated voltage range (± 10 %): ....................................................................................................................... 100 V 
 240 V
Rated frequency (± 10 %): ..................................................................................................................................... 50/60 Hz
Power consumption: ................................................................................................................................................... 35 VA
Protection class: .................................................................................................................................................................. I
Power fuses: .................................................................................................................................................... T 1 A L 250 V
Pollution degree: ................................................................................................................................................................. 2
Overvoltage category: ........................................................................................................................................................ II
Electrical overload protection: ..................................................................................................................................... yes
Internal current limit for the electronics: .................................................................................................................... yes
Dimensions:
L x W x H: ................................................................................................................................ 600 mm x 250 mm x 230 mm
Height with magnifier support ............................................................................................. 600 mm x 250 mm x 320 mm
Height with microscope: ....................................................................................................... 600 mm x 250 mm x 490 mm
L x W x H control unit (when the bases are folded in): ...................................................... 165 mm x 120 mm x 72 mm
Weight:
Without magnifier support and control unit ............................................................................................................ 56 kg
VT1200 control unit ........................................................................................................................................................ 1 kg
Magnifier support .......................................................................................................................................................... 2 kg
Microscope support with stereomicroscope ............................................................................................................ 4 kg