Installing the Substage Condenser
Substage Condenser
 
The Leica DM750 has an open substage 
condenser mount, therefore the conden-
ser needs to be installed.
1.  Move the specimen stage upwards as far as 
it will go by using the coarse focusing knob 
on the side of the microscope.
2.  Move the condenser holder into the lowest 
position by using the condenser focusing 
knob on the left side of the stage mount.
3.  Unscrew the two setscrews (or wings-
crews for Koehler stands) on the condenser 
holder. 
 
The substage condenser mount is open