The  ring  illuminator,  the  Leica  LED5000  RL 
("Ring Light"), is installed on the objective using 
a single screw. It has been optimized for a work-
ing distance between 60 and 70 mm.
Constraints
The Leica LED5000 RL can be used in conjunc-
tion with the planapochromat 1× and planapo-
chromat  0.63×  objectives.  With  all  other 
objectives, the working distance is too low for 
adequate illumination.
The  ring  illuminator  cannot  be  used  together 
with the objective nosepiece.
Assembly
• 
1.  Connect  the  CAN-bus  cable  to  the 
ring illuminator. The flat part of the plug 
must be facing upwards.
• 
2.  Push  the  ring  illuminator  over  the 
objective as far as it will go and screw it 
into place.
• 
3.  Plug the other end of the cable into 
one  of  the  two  "CTL2"  sockets  on  the 
column.
For  optimum  accessibility  of  the  specimen, 
the  ring  illuminator  should  be  installed  with 
the cable facing backwards. However, it is also 
possible to turn the ring illuminator sideways, 
for  example  if  simultaneously  using  the  Leica 
LED5000  MCI™  system  illumination.  In  this 
case, the ring illuminator cannot be connected 
directly to the Leica LED5000 MCI™.
Leica LED5000 RL