Leica M205 C Manuel D’Utilisation

Page de 114
Leica M series 
Manual 
60
All  Leica  eyepieces  are  also  available  with 
built-in dioptric correction, allowing the micro-
scope to be used without glasses even by those 
with vision problems. The correction comprises 
±5 diopter settings.
Using the Dioptric Correction
1.  Set the dioptric correction of both eyepieces 
to the mid position ("0" diopter settings.
2.  While  wearing  your  glasses,  look  through 
the eyepieces and focus on the specimen.
3.  Rotate  both  eyepieces  to  the  maximum 
value of "+5".
4.  Hold  one  eye  closed  and  rotate  the  other 
eyepiece in "-" direction until the specimen 
appears sharp.
5.  Then,  open  the  other  eye  and  correct  the 
diopter settings until the image is uniformly 
sharp.
 
Note that when using dioptric correction, 
the advantage of parfocality is lost—thus 
you  have  to  manually  refocus  each  time  you 
change the zoom level. To also use parfocality 
with  dioptric  correction,  refer  to  the  instruc-
tions on page 61.
Dioptric correction