Image acquisition 
3.1
 
Accessing the computer 
If  the  computer  is  turned  off,  push  the  power  button  to  turn  it  on.  When  the  login 
screen appears, log in using your hyad login account. Check that the domain is set to 
LTDK.  
3.2
 
View Finder application 
A 10-bit Olympus DP50 color CCD camera is attached on the top of the microscope. 
It is controlled by a computer and Studio Lite software. 
To  capture  an  image,  open  the  Pixera  Corporation  Studio  Lite  program  by  double-
clicking its icon on the  desktop (Figure 14). Then, open up View Finder  by  clicking 
the camera button in the menu at the top. 
    
     
 
 
Figure 14. Starting the Studio Lite and View Finder applications 
 
 
Figure 15. Adjustments in View finder window 
 
Press  ‘Live  image’  button  (green  triangle  in  Figure  15)  in  the  View  Finder 
window. A live image of your sample will be shown on the screen. The speed 
of  the  image  refresh  can  be  adjusted  by  selecting  Fast,  Zoom,  or  Full  in  the 
main  menu  under  View  Finder  Mode.  It  is  easier  to  focus  and  move  the 
sample  with  the  Fast  setting.  For  fine  adjustment  of  the  focal  plane,  use 
Zoom
3.3
 
Adjustments before image acquisition 
3.3.1
 
Capture resolution 
There  are  three  resolutions  (640x480,  1392x1040,  and  2776x2074)  to  choose  from 
(Figure  15).  Use  higher  resolution  for  more  details,  if  the  bigger  file  size  does  not 
matter.  The  highest  resolution  is  slower  to  acquire  than  the  other  two.  Resolution 
selection does not affect the live view image.  
3.3.2
 
Sensitivity 
In  the  Sensitivity  box,  select  ISO  values  from  50  to  400.  Higher  value  shortens  the 
exposure time but increases background noise and decreases image quality. In digital 
imaging,  ISO  100  is  the  level  of  intensity  read  from  the  sensor.  Everything  else  is