Leica B523 Manuel D’Utilisation

Page de 18
 
 
15 
 
Cover the microscope with the red hood 
6
 
Troubleshooting 
6.1
 
There is no light 
 
Microscope is not turned on (see Section 2.1) 
 
Light intensity set too low (see 2.1 on page 1) 
 
Do you see any light in the lamp chamber? If not, the halogen bulb may need 
to be changed - contact MIU 
6.2
 
Light is too dark 
Check:  
 
whether the polarizer or other filters are on the light path (Section 2.9 on page 
8) 
 
that aperture diaphragm is not too closed (Section 2.8 on page 6) 
6.3
 
Problems with focus or picture clarity 
Check that: 
 
no filters are on the light pathway 
 
Koehler illumination is correctly done (Section 2.8 on page 6) 
 
objective is not dirty 
 
oculars are not dirty 
 
oculars are correctly adjusted (Section 2.6) 
If image quality does not satisfy you, you should take your images with AxioPlan 
microscope, which has better optics and camera 
6.4
 
View is partial 
Check that:  
 
none of the filters are partially on the light path 
 
the polarizer is not partially on the light path 
 
the objective is correctly in place 
 
the field diaphragm is not too closed 
6.5
 
Colors have red/yellow tint 
 
Check the white balance 
 
For other color errors, check the color filters (Section 2.10) 
6.6
 
Ocular view is not at the same focus as the camera view 
 
Check ocular adjustment (Section 2.6)