Leica RM2125 Manuel D’Utilisation

Page de 15
Leica RM2125 Troubleshooting Guide
Problem
10. Microtome makes "clunking noise" when sectioning
Cause(s) / 
Corrective Action
I)   Linear bearings in cage are at their lower limits.
a) Linear bearings need to be cleaned and possibly re-greased.
II)   Alignment of bearing cage is incorrect.
b) Schedule a service call.
13