Cleaning and Maintenance
• 
Before  carrying  out  cleaning  tasks, 
switch  off  the  Leica  EL6000  compact 
light source. 
Use only a dry, lint-free cloth for cleaning. When 
cleaning, ensure that no dust or other foreign 
bodies enter the device through the ventilation 
slots. 
In  case  of  damage  (even  external  damage), 
immediately  stop  using  the  Leica  EL6000 
compact light source and send it to service.
If it will not be used for a long time, protect the 
Leica  EL6000  compact  light  source  from  dust 
using a suitable cover (such as the plastic bag 
from the original packaging).
Spare parts and accessories
The  only  parts  of  the  Leica  EL6000  compact 
light source that may be replaced by the user 
are the instrument fuses and the lamp. 
The  replacement  of  the  lamp  is  described 
 
on Page 61.
Replacing the instrument fuse
1.  Disconnect  the  Leica  EL6000  from  the 
power supply.
2.  Open the drawer with the instrument fuses 
located close to the power supply input.
3.  Replace the instrument fuse. 
Repairs
Opening  or  working  on  the  Leica  EL6000 
compact light source in any manner other than 
described  above  (changing  lamps,  changing 
fuses) may be carried out by the supplier of the 
instrument  or  the  supplier's  authorized  repre-
sentatives.
Permitted replacement fuses
Fuse: 5×20, 2.5 A, slow-blowing, breaking capacity H
Wickmann Series 181 
www.wickmann.com
Schurter Series SPT 5×20   www.schurter.ch