121
Observation without Fluorescence
 
We recommend that you first familiarize 
yourself  with  the  controls  of  your  Leica 
fluorescence  microscope  in  transmitted  light 
without the presence of fluorescence illumina-
tion.
• 
After  switching  on  the  high-pressure 
mercury burner, you must adjust the arc 
lamp immediately (see User Manual M2-216-1).
 
We recommend that you first familiarize 
yourself with the basic functions of your 
Leica stereomicroscope as they are described in 
the previous section of this User Manual. These 
include the binocular tube, the eyepieces, the 
zoom  magnification  changer,  the  focusing 
drive, and others.
The  following  describes  the  specific  functions 
of the fluorescence system only. 
Observation without Fluorescence
The FLUOIII® filter system includes a UV shutter 
that can be used to close the illumination beam 
path.  This  feature  protects  the  mercury  lamp, 
which  should  not  be  switched  on  and  off  too 
frequently.  If,  for  instance,  you  need  to  inter-
rupt your work, you do not have to switch off 
the  high-pressure  mercury  burner  to  preserve 
sensitive specimens from being burned out or 
bleached out. 
 
Switching  the  high-pressure  mercury 
burner  on  and  off  frequently  decreases 
its life and causes delays, as the lamp can only 
be switched back on after it has cooled. Instead, 
you  can  close  the  UV  shutter  and  reopen  it 
when you return to your workplace.
Close the UV shutter when you do not want to 
view the specimen with fluorescence illumina-
tion, but only with transmitted light.