Leica M205 FA Manuel D’Utilisation

Page de 152
Leica M-series 
User Manual 
150
Calculating the Total Magnification and Field of View Diameter
Parameter
MO
Magnification of the objective 
ME
Magnification of the eyepiece 
z
Magnification of the changer position
q
Tube  factor,  e.g.  1.5×  for  coaxial  incident  light,  1.6×  for  45° 
ErgoTube™
r
Factor  1.25x  if  the  planachromatic  and  planapochromatic 
objectives  of  the  MZ125/MZ16  are  used  on  the  MS5,  MZ6, 
MZ75 or MZ95
NFOV Field number of the eyepiece. Field numbers are printed on the 
eyepieces: 10× = 21, 16× = 14, 25× = 9.5, 40× = 6.
Example 
MO
1× objective
ME
25×/9.5 eyepiece
z
Zoom position 4 
q
Coaxial reflected light 1.5×, tube factor
r
Factor 1.25×
Calculation example: Magnification in the binocular tube:
MTOT VIS = MO × ME × z × q × r 
or  
1 × 25 × 4 × 1.5 × 1.25 = 187.5× 
Calculation example: Field of view diameter in the specimen
∅ OF:
NFOV
MO × z ×q ×r