Preparations
• 
Hold  the  objectives  firmly  during 
assembly and disassembly so that they 
do not fall onto the stage plate.
Move the drive housing all the way upwards 
 
ρ
and remove the optics carrier, if the carrier 
has already been installed.
Assembly
1.  Remove  the  transport  anchor  from  the 
objective nosepiece.
2.  Rotate the moving part by 90° and attach 
the objective nosepiece to the drive hous-
ing  from  the  front  (!).  Screw  the  objective 
nosepiece firmly into place.
3.  Unscrew  the  three  Phillips  screws  on  the 
objective  mount  of  the  optics  carrier  and 
remove the intermediate ring.
4.  Screw the optics carrier onto the objective 
nosepiece.
5.  Screw  both  objectives  onto  the  objective 
nosepiece.  It  makes  no  difference  which 
position an objective occupies.
6.  Unscrew the locking screws on both sides 
of the objective nosepiece.
You can now adjust the parfocality (see instruc-
tions on the next page).
Objective Nosepiece – Assembly