Setting up the instrument
• 
Set up the Leica EL6000 so that the front 
is readily accessible and visible.
The ventilation slots on the sides and back 
 
ρ
panel of the device must not be covered.
Maintain a free room of at least 150 mm on 
 
ρ
the rear panel of the instrument.
Instructions for lamp replacement
• 
The  lamp  contains  mercury!  Be  abso-
lutely  certain  to  follow  the  handling 
instructions and safety notes provided with the 
lamp.
• 
Maintain cleanliness when inserting the 
lamp.  Remove  any  clinging  packaging 
materials.
• 
The  lamp  used  in  the  instrument 
becomes very hot during operation and 
has  high  internal  pressure  when  hot.  Before 
changing the lamp, it is mandatory to allow the 
instrument to cool off for at least 20 minutes.
• 
The Leica EL6000 compact light source 
comes  with  a  factory-supplied  lamp, 
but the lamp is not installed when shipped. This 
minimizes  the  danger  of  damage  to  the  lamp 
during transport.
• 
Before  transporting  the  Leica  EL6000 
compact light source, you first have to 
remove the lamp.
Inserting the lamp
1.  Pull  the  power  plug  out  of  the  socket,  so 
that  the  EL6000  is  disconnected  from  the 
power system.
2.  Unscrew both screws of the cover using a 
3 mm Allen key.
3.  Remove the housing cover.
Leica EL6000 – Preparations