Leica M205 FA Manuel D’Utilisation

Page de 152
Leica M-series 
User Manual 
78
Adjusting
1.  Set  the  dioptric  correction  for  both 
eyepieces to "0".
2.  Select  the  lowest  magnification  and  focus 
on a flat specimen.
3.  Select the highest magnification and read-
just the sharpness.
4.  Select the lowest magnification again, but 
do not look into the eyepieces. 
6.  Rotate  the  eyepieces  counterclockwise 
in  the  "+"  direction  as  far  as  they  will  go  
(+5 diopter settings).
Leica stereomicroscopes are parfocally matched. 
The prerequisite for this is the correct setting of 
the diopters and the parfocality. The following 
adjustments  only  have  to  be  carried  out  once 
by each user.
Preparations
Move the lever of the video/phototube to 
 
ρ
the  "observation"  position  and  open  the 
diaphragm.
If you are using the microscope carrier AX, 
 
ρ
set it to stereoscopic observation.
7.  Look into the eyepieces.
8.  Slowly  rotate  each  eyepiece  individually 
in the "–" direction until each eye sees the 
object sharply imaged. 
9.  Select  the  highest  magnification  and 
 refocus if necessary.
Now, if you adjust the magnification from the 
lowest  to  the  highest  level,  the  specimen  is 
always brought into sharp focus. If not, repeat 
the process.
Dioptric Correction and Parfocality