Leica DMI6000B Manuel D’Utilisation

Page de 232
97
8. Operation
 
Note:
When replacing objectives, you must perform a
teach-in for the new objectives in the Leica Ap-
plication Suite (LAS). A parfocality teach-in
should also be performed.
Note:
For lockable immersion objectives lock these by
pushing the front part upwards until it stops
(approx. 2 mm). Then, after a gentle turning mo-
tion to the right, the objective is locked.
For objectives with corrective mounts turn the
knurl to adjust the objective to the thickness of
the cover glass.
Color coding of objectives
The magnification of each objective is indicated by a color ring in accordance with DIN/ISO stand-
ards:
Immersion objectives are marked by an addi-
tional, lower color ring.
black
oil or Imm (universal objective for
oil, water or glycerin)
white
water
orange
glycerin
The various engraved markings of the objectives
provide information on their applications:
black or
bright field objectives,
dark blue
strain-free
green
phase contrast objectives,
strain-free
      100x             63x             40x           25x            16x            10x           6.3x            4x           2.5x            1.6x
      125x                                50x           32x            20x                                                5x
      150x
      160x
     white          dark-          light-        dark-        light-       yellow      orange        red       brown         gray
                          blue            blue        green       green