Leica M205 A Manuel D’Utilisation

Page de 114
Leica M series 
Manual 
26
Objective Nosepiece – Adjusting Parfocality
The following procedure only has to be carried 
out once. Afterwards, both objectives are parfo-
cal, meaning that the specimen remains in focus 
when the objective nosepiece is rotated.
This procedure must be repeated if you replace 
either of the two objectives with another.
The  following  example  assumes  the  combi-
nation  of  the  1×  and  2×  planapochromats.  If 
you  are  using  another  objective  combination, 
replace the 2× objective in the description with 
the objective with the stronger magnification.
Preparation
Open the iris diaphragm.
 
Set the dioptric correction of the eyepieces 
 
to "0".
Adjustment
1.  Rotate the 2× objective into the beam path 
and set it to the lowest magnification.
2.  Focus on the specimen.
3.  Rotate  the  1×  objective  into  the  beam 
path.
4.  Turn  the  objective  on  the  thread  in  both 
directions  until  the  specimen  appears 
sharp.
5.  Toggle to the 2× objective.
6.  Select  the  strongest  magnification  and 
refocus  until  the  specimen  appears  abso-
lutely sharp.
7.  Toggle to the 1× objective.
8.  Turn  the  objective  on  the  thread  in  both 
directions  until  the  specimen  appears 
absolutely sharp.
By means of zooming, check that the behavior 
of  the  objective  is  parfocal.  Repeat  the  check 
with  the  other  objective.  If  it  is  not  parfocal, 
repeat the procedure.
9.  Tighten the locking screws.