Maintenance, Calibration and Troubleshooting
Maintenance
4
 
4-1 
Section 4 
Maintenance, Calibration and Troubleshooting
 
 
4.1 Maintenance 
The 943-227-15 Digital Preamplifier is designed to operate for extended periods of time with no 
scheduled maintenance required.  Operation may be verified by periodically actuating the check source 
(provided with the detector and sampling assembly) and observing the response of the unit.  If the 
response varies by more than 50% of the normal value, further troubleshooting may be required.  If a 
problem develops, troubleshoot the unit per Section 4.3 and the drawings in Appendix B.  
 
Whenever the preamplifier enclosure is opened, the condition of the internal desiccant package should be 
checked.  The desiccant bag is provided in the enclosure to absorb moisture.  The desiccant bags should 
be checked for moisture absorption prior to placing into service, or whenever the electronics enclosure 
cover is opened.  Each bag contains a color indicator stripe that turns red when it should be replaced.
 
 
 
4.2 Electronic Calibration 
Electronic Adjustments 
 
There are no user adjustments available on the 943-227-15 Digital Preamplifier.  The Programmable Gain 
Amplifier and Auto Zero are factory adjustments, requiring the use of a factory test PROM and test fixture.  
We strongly recommend the unit be returned to the factory for repair or adjustment. 
 
Table 4-2 lists the electronic adjustments for the 943-227-15 digital preamplifier.  The procedures for 
making these adjustments are contained in the factory Calibration Procedure listed below.  Note that 
special test firmware is required to perform the electrical adjustments. 
   
Table 4-1.  Electronic Adjustments 
 Potentiometer 
Adjustment 
(943-227-15) 
Tolerance 
 
R11 
Programmable Gain Amplifier Offset Adjustment 
± 2 mV 
 
R12 
Auto Zero Set-Up Adjustment 
± 10 mV 
 
 
 
NOTE 
Refer to the drawings in Appendix B to locate adjustment 
and/or test points, if required.
 
 
 
The follow factory calibration and test procedures are listed for reference only. 
 
Document 
Description 
 
TP943-227-15VL Test 
Procedure, 
Victoreen Loop Communications 
TP943-227-15 
Test Procedure, RS232 communications 
TP977-200-15M 
Electrometer PCB Test Procedure