Emerson HAS8xE-IM-ATEX Manuel D’Utilisation

Page de 54
MLT 2 / BINOS
 100 F
Emerson Process Management Manufacturing GmbH & Co. OHG
1 - 4
ATEX Instruction Manual
HAS8xE-IM-ATEX
January 2004
1-3-4    Purge Gas Conditions
1-3-4
Purge Gas Conditions
Purge gas:
•  Inert gas (e.g. Nitrogen)
•  Air (from an ex-free zone)
Temperature:
As ambient, but min. 20 °C to 35 °C.
Input pressure at
inlet of EEx p device:
•  3,000 to 4,000 hPa (3 to 4 bar)
Minimum internal
overpressure of
the enclosure:
•  1 mbar
Maximum internal
overpressure of
the enclosure:
•  25 mbar
Pre-purging phase
Single enclosure analyzer
Dual enclosure analyzer
Duration of
pre-purging phase:
5 min.
17 min.
Flow rate during
pre-purging phase:
8 Nm³/h
7 Nm³/h
(= 133.4 l/min.)
(= 117 l/min.)
Flow rate during
pre-purging phase
22 Nm³/h
22 Nm³/h
(@ 25 mbar overpressure):
(= 367 l/min.)
(= 367 l/min.)
Medium has to be dry and free from
dust, oil, corrosive or aggressive
components!