Emerson Process Management NGA2000 HFID Manuale Utente

Pagina di 76
Instruction Manual 
748297-E 
June 2003 
Rosemount Analytical Inc.    A Division of Emerson Process Management  
Maintenance  and Troubleshooting     4-3 
Model NGA2000 HFID 
Oven Assembly 
659551
4-2 COMPONENT 
REPLACEMENT 
a. Oven 
Though the oven can be replaced as a 
complete unit, all internal components are 
field replaceable.    
Oven Removal 
Refer to Figure 4-3 below, disconnect the 
oven's three gas lines and seven electri-
cal cables, noting location of mating con-
nectors for re-installation. 
Note 
 DO NOT remove the fittings from the 
gas lines on the detector. 
Remove the two hex nuts securing the 
oven to the chassis and the two screws 
securing oven to the rear panel.  Lift oven 
assembly from analyzer.
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Figure 4-3. 