Performance Procedures
4-24
Testing the Monitor
3
Set the temperature simulator as follows:
a. Temperature: 37
°
C (98.6
°
F).
b. Probe Type: YSI 400 Series
4
Press the On/Standby button to turn the monitor On.
5
After the power-up sequence, verify the temperature reads 37°C ±0.1°C (98.6°F
±0.2°F if Fahrenheit is selected as the temperature unit).
±0.2°F if Fahrenheit is selected as the temperature unit).
6
Turn the monitor Off.
Step
Action