Performance Procedures
4-38 
Testing the Monitor  
ECG Sync 
Test
This test checks the performance of ECG synchronization between the monitor and a 
defibrillator. The ECG sync performance test is required once every year and when the 
monitor is repaired or when the monitor’s parts are replaced.
Tools Needed:
Dynatech Nevada MedSim 300 or equivalent simulator
Defib Sync Cable (M4820A) or Switchcraft 850
Switchcraft 750 
Two sections are needed to complete the connection from the monitor to the simulator.
1. If a Defib Sync Cable is available, connect it the ECG Sync output on the monitor, then 
to the Switchcraft 750. 
or 
If a Switchcraft 850 is available, connect it the ECG Sync output on the monitor, then 
to the Switchcraft 750.
2. Connect the Switchcraft 750 to the SYNC/A PACE input on the simulator
Connect to the monitor
Connect to the simulator
Defib Sync Cable
Switchcraft 750
Switchcraft 850
Switchcraft 750
Step
Action
1
Connect the ECG leads and trunk cables between the monitor and the simulator 
(as described above).
2
Turn on the monitor and the simulator
3
On the simulator main menu select DEFB, then CARD to set the simulator to 
Defibrillator Tests/Cardioversion. 
4
Press Start to begin the test. If the monitor is working properly, the following 
sequence of text messages will appear on the simulator over a period of 
approximately 3 seconds: "cardioversion: afib", "cardioversion: sync ok", and 
"cardioversion: converted" (this message will persist on the simulator)
5
If the monitor isn’t working properly, messages such as the following may 
appear: "cardioversion: vfib", "cardioversion: late".