Calculating Temp Difference
The monitor can calculate and display the difference between two temperature values by subtracting 
the second value from the first. The difference is labeled 
Temp
1
In the 
Setup Main
 menu, select 
Measurements
.
2
In the 
Setup
 
Temp 
menu, select
 First Temp
.
3
Label the measurement source as appropriate.
4
Select 
Second Temp
.
5
Label the measurement source as appropriate.
Temp
non-specific temperature label.
Trect
rectal temperature
Tart
arterial temperature
Tskin
skin temperature
Tcore
core temperature
Tven
venous temperature
Tesoph
esophageal temperature
Tnaso
nasopharyngeal temperature
T1, T2, T3, T4
Non-specific temperature labels
Tamb
ambient temperature
Tcereb
cerebral temperature
Ttymp
tympanic temperature
Tvesic
vesical temperature