15 Monitoring Airway Flow, Volume and Pressure
Setting up Spirometry
194
3
Select the appropriate setting.
Changing the Temperature of the Inspired Gas
Depending on the type of ventilator environment (e.g. if you are using a Heat Moisture Exchanger) 
you may want to change the temperature setting of the inspired gas:
1
In the 
Setup Spirometry 
menu, select 
Gas Compensation
.
2
Select 
Inspired Temp 
to change the temperature of the inspired gas.
3
Select the appropriate temperature setting.
N O T E
Changing the temperature setting unintentionally will result in approximately 3% reduced accuracy 
per 10°C ( 18°F) temperature setting change.