Philips MP60/70/80/90 Manuale Utente
239
22
22
High Resolution Trend
Waves
High resolution (hi-res) trend waves provide information for fast-changing parameters at a glance.
They can help you to recognize trend patterns and to find interrelationships between measurements, by
comparing changes in your patient’s condition across different hi-res trend waves. This can aid in the
early detection of cardiorespiratory and hemodynamic problems.
They can help you to recognize trend patterns and to find interrelationships between measurements, by
comparing changes in your patient’s condition across different hi-res trend waves. This can aid in the
early detection of cardiorespiratory and hemodynamic problems.
The amount of data you can show on a screen depends on the display size and resolution: for example,
on an eight inch (20 cm) wave trace on your monitor screen, you can view approximately six minutes
of hi-res trend wave trace. The hi-res waves are drawn at a speed of 3 cm/minute.
on an eight inch (20 cm) wave trace on your monitor screen, you can view approximately six minutes
of hi-res trend wave trace. The hi-res waves are drawn at a speed of 3 cm/minute.
The availability of hi-res trend waves depends on your monitor configuration and on the options
purchased for your monitor.
purchased for your monitor.
Changing the Hi-Res Trend Waves Displayed
♦
To change the selection of hi-res trend waves on the Screen, select the hi-res trend wave and select
the required hi-res trend wave from the list of available measurements.
the required hi-res trend wave from the list of available measurements.
If there are no hi-res trend waves on the Screen, you must change to a Screen with hi-res trend waves.
Hi-Res Trend Wave Scales
High-resolution trend waves are either compressed realtime waves or fast numeric trends.
• Hi-res trend waves from compressed realtime waves include: Resp, CO
2
, ABP, PAP, CVP, ICP, AWP,
anesthetic agents.
• Hi-res trend waves from fast numeric trends include: btbHR, SpO
2
, tcpO
2
, tcpCO
2
, Pulse, Perf,
CPP, BIS, CCO, inO
2
, Delta SpO
2
.
Hi-res trend waves from waves use the wave scale as their display scale. Changing the wave scale
changes the hi-res trend wave scale.
changes the hi-res trend wave scale.
Hi-res trend waves from numerics use a pre-defined display scale dependent on the patient category.
Hi-Res Trend Waves and OxyCRG
OxyCRG is a combination of three hi-res trend waves, btbHR, SpO
2
, and Resp, mainly used in
neonatal monitoring.