Changing Loops Display Size
29 Respiratory Loops
291
Changing Loops Display Size
To change the loops display size, in the 
Loops
 window, select the pop-up key 
Select Size
 then 
select 
Size x 0.5
 to display loops at half the usual size
Size x 1
 to display loops at the usual size
Size x 2
 to display loops at twice the usual size
This does not affect the size of loops printed in reports.
Using the Loops Cursor
To view the coordinates for any point on a stored loop, 
1
Select the pop-up key 
Select Loop
.
2
Select the required loop from the list of available loops.
3
Use the arrow pop-up keys to move the cursor around the loop. The values for both loop axes are 
shown for every point on the loop. 
Select the X at the top of the loop list to exit the cursor. 
Changing Loops Type
To change the loop type, in the 
Loops
 window, select the 
Loop Type
 pop-up key (only available if 
the source device delivers three waves), then select
Press/Volume
 to display Pressure/Volume loops
Flow/Volume 
to display Volume/Flow loops
Press/Flow
 to display Pressure/Flow loops.
or, if the source device only delivers two waves,
1
In the 
Loops
 window, select the 
Setup Device
 pop-up key to enter the 
Setup <VueLink Device>
 menu.
2
Select 
Wave 1
 and select pressure, flow, or volume waves as required. Repeat for 
Wave 2
.
Setting Up Source Device
The source device for the loops is indicated in the 
Loops
 window title. 
Select the 
Setup Device
 pop-up key to view the settings of the current source device. 
If the source device is changed, all reference loops from a previous device are cleared from the reference 
list when the first loop from the new source device is captured. Reconnecting the previous device recalls 
its reference list.