Philips MP40/50 Manuale Utente

Pagina di 414
32 Maintenance and Troubleshooting
Maintenance Task and Test Schedule
306
4
Inspect the patient cables, leads and their strain reliefs for general condition. Make sure there are 
no breaks in the insulation. Make sure that the connectors are properly engaged at each end to 
prevent rotation or other strain.
5
Apply the transducer or electrodes to the patient, and with the monitor switched on, flex the 
patient cables near each end to make sure that there are no intermittent faults.
Maintenance Task and Test Schedule
The following tasks are for Philips-qualified service professionals only. All maintenance tasks and 
performance tests are documented in detail in the service documentation supplied on the monitor 
documentation CD. 
Ensure that these tasks are carried out as indicated by the monitor’s maintenance schedule, or as 
specified by local laws. Contact a Philips-qualified service provider if your monitor needs a safety or 
performance test. Clean and disinfect equipment to decontaminate it before testing or maintaining it.
Maintenance and Test Schedule
Frequency
Monitor Tests
Safety checks. Selected tests on the basis 
of IEC 60601-1
At least once every two years, or as needed, after any repairs 
where the power supply is removed or replaced, or if the 
monitor has been dropped.
Monitor Maintenance
Check ECG synchronization of the 
monitor and defibrillator (only if 
hospital protocol requires use of 
monitor during defibrillation)
At least once every two years, or as needed. 
Replace backlight 
(integrated displays only)
25,000 - 30,000 hours (about three years) of continuous usage, 
or as needed.
Measurement Server and Module Tests
Performance assurance for all 
measurements not listed below.
At least once every two years, or if you suspect the measurement 
values are incorrect.
Measurement Server and Module Maintenance
NBP calibration 
At least once every two years, or as specified by local laws.
Microstream CO
2
 calibration and 
performance test
At least once a year or after 4000 operating hours.
Mainstream and sidestream CO
2
 
calibration check (M3014A)
At least once a year, or if you suspect the measurement values are 
incorrect.
Battery Maintenance
Battery
See the section on Maintaining Batteries