Using the C.O. Procedure Window
13 Monitoring Cardiac Output
163
Using the C.O. Procedure Window
The C.O. procedure window displays up to six trials (measurement curves) with the trial number and
the C.O. value under the thermodilution curve. When you open the window, a line of pop-up keys
automatically appears to let you carry out C.O.-related tasks. This example shows the procedure
window for the transpulmonary (PiCCO) Method. The window may be configured to look slightly
different on your monitor.
the C.O. value under the thermodilution curve. When you open the window, a line of pop-up keys
automatically appears to let you carry out C.O.-related tasks. This example shows the procedure
window for the transpulmonary (PiCCO) Method. The window may be configured to look slightly
different on your monitor.
To open the C.O. procedure window,
♦
Select
Cardiac Output
in the
Setup C.O.
or
Setup CCO
menu, or
♦
Select the
Cardiac Output
SmartKey on the Screen, if configured, or
♦
Press the
START
hardkey on the front of the C.O. plug-in module, if available, or
♦
Press a remote start switch, if you are using one.
♦
To change the measurement parameters shown in the results table of the C.O. procedure window,
select the
select the
Table Contents
pop-up key and choose from the list of available parameters.
♦
To view the currently-used temperature unit, see the “grayed-out” setting
Temperature Unit
in the
Setup C.O.
menu. This setting can only be changed in Configuration Mode.
♦
The Cardiac Output screen element may be configured to display permanently on a specially
designed Screen. Selecting the screen element opens the cardiac output pop-up keys.
designed Screen. Selecting the screen element opens the cardiac output pop-up keys.
Trial curves
Trial curve scale
Averaged values
Results table of current trial
Prompt
message field
message field
Thermodilution
curve of current
trial
curve of current
trial
Curve alert
message field
message field
Setup
information
information
Current trial
number
number
Continuously
measured
numeric
measured
numeric
Cardiac Output (Transpulmonary)
CathCt
InjVol
BSA 1.93
C.I. 4.40
C.O. 8.49
Average
Unit
C.O.
CFI
C.I
ITBVI
ETVI
Tinj
InjVol
Tblood
Tblood 37.0
l/min
l/min/m
2
ml/m
2
o
C
ml
#1
Cal
Cal
#3
Cal
Cal
#2
Cal
Cal
0.3
o
C
Trial
CCO-CAL
8.51
5.5
4.41
996
4.41
996
8.53
5.5
4.42
996
6.0
2.0
10
11.4
6.0
5.88
1225
5.0
2.0
10
8.49
5.5
4.40
998
6.1
2.0
10
Ready for new measurement
Start
C.O.
Stop
C.O.
Select
Trial
Accept/
Reject
Save C.O.&
Cal CCO
Print/
Record
Table
Contents
Hemo
Calc
Change
Scale
Setup
C.O.