Introduction 
 
 
 
Teledyne Analytical Instruments 
13
 
 
Introduction 
1.1 Overview 
Teledyne’s Model LXT-220 Dissolved Oxygen Controller is a 
robust and versatile instrument for the measurement and process control 
of dissolved oxygen in a liquid component. Highly customizable, the 
Model LXT-220 comes standard with a single 4-20mA current output 
but can have as many as five. The current output is field selectable and 
may be expanded to any range within the transmitter’s full operating 
range. It can also be programmed as reverse acting; for example, a 4-20 
mA output for a range of 0 to 10 ppm can be reversed to 10 to 0 ppm. 
The standard instrument accepts a single input typically from a 
Teledyne SP1 or SP2 Dissolved Oxygen probe however a dual input 
option is available. This optional dual input extends the functionality of 
the instrument to accept conditioned input from a variety of sensors for 
specific process measurements such as: 
•  Second Dissolved Oxygen (DO) channel 
•  Resistivity/conductivity 
•  pH  
•  Oxidation-Reduction Potential (ORP) 
•  Specific Ion  
For process control, the microprocessor based LXT-220 
Transmitter employs a multi-bus system architecture for PID 
temperature control and sensor-specific process variable outputs such as 
DO, pH, ORP, Resistivity/Conductivity, and ion-specific element (ISE) 
output as a function of multiple inputs. Up to six field-configurable 
alarm or control relay outputs are available to the user to match 
individual process needs. These relays can be field configured as high or 
low alarms. Hysterisis (dead band) is defined by setting the ON and OFF 
relay values. These relays can be configured to any of the input 
parameters. (See Section 3.3.5).