Dissolved Oxygen Controller  
Operation 
 
 
 
Teledyne Analytical Instruments 
33
 
 
The protection level for the operational displays is the number on 
the left side of the second line. None of the operational displays are 
password protected if the protection level is 0. If it is not zero then, with 
the exceptions shown in Table 3-1, all displays that affect the operation 
of the instrument are protected. An N entry in the table means that the 
display is not protected. 
 
Table 3-1: Protection Level Exceptions 
level 
contrast 
man out 
1-point buf 
3.3.1.1 C
HANGING 
P
ROTECTION 
L
EVEL
 
To change the protection level, navigate using the right side 
(CALIBRATE) ◄/ ► keys to the old protection level number (3 in the 
above screen). Adjust the numerical value using the right side 
(CALIBRATE) ▲/▼ keys and then accept the input by pressing either 
of the left side MENU SELECT) ▲/▼ keys. The old operational 
password is erased if the protection level is changed to 0. Changing from 
0 to another value has no effect until a new password is entered. The 
display shows no PW if the password has been erased. 
3.3.1.2 C
HANGING 
O
PERATIONAL 
P
ASSWORD
 
To enter a new operational password, navigate to the old password 
(or no PW) using the (CALIBRATE) ◄/ ► keys and select by pressing 
both (CALIBRATE) ◄/ ► keys simultaneously. This brings up the new 
operational password screen. 
 
New opr password