Teledyne LXT-220 Manuale Utente

Pagina di 61
Maintenance 
LXT-220 DO 
 
 
Teledyne Analytical Instruments
 
52 
 
Table 4-1: Troubleshooting  
SYMPTOM 
POSSIBLE CAUSES 
SUGGESTED 
ACTION 
No power to the 
instrument. 
Check power supply 
to be sure the correct 
voltage is being 
supplied to the 
transmitter. 
Contrast level is set 
too low. 
Set contrast level to a 
higher value.  
Because the contrast 
menu is always the 
top menu, it can be 
reached by pressing 
the upper MENU 
SELECTION key at 
least 12 times.  Enter 
the calibrate mode 
and increase the 
value. 
Sensor or signal 
conditioner has a 
short that draws too 
much power from the 
transmitter. 
Check the sensor for 
proper operation.  To 
verify, disconnect all 
sensor wires and 
cycle power off, then 
on. 
The LCD does not 
display. 
The PROM has not 
been properly 
installed in the socket. 
Check orientation of 
the PROM to the 
socket. Make sure the 
pins are fully inserted 
and are not bent. 
Readings are not 
linear at the low end 
of the range. 
The sensor range is 
not compatible with 
the instrument. 
Verify the sensor 
range.  Install a 
sensor with the 
correct range. 
Erratic readings. 
Air bubble in sensor. 
Check orientation of 
sensor.