SECTION 6—CYLINDER FILL TIMES
20
SECTION 6—CYLINDER FILL TIMES
NOTE: All filling times are approximate and may vary on 
environmental conditions.
*
Troubleshooting
If post valve cylinder fails to function, try the following 
solutions. If problem cannot be corrected, consult your 
dealer/home care equipment provider.
Replacement Parts - Accessories
HF2POST6BAG or HF2POST9BAG Carrying Bag.
Cylinder 
Filling 
Times
Concentrator Flow Rate to Patient:
(Platinum 5 up to 3 L/min. and Platinum 10 up to 5 
L/min)
HF2POST6
HF2POST9
HF2POSTD
NORMAL 
1 hr 25 min
2 hr 20 min
3 hr 20 min
MAXIMUM
1 hr 50 min
2 hr 45 min
3 hr 50 min
PROBLEM
POSSIBLE CAUSE
SOLUTION
No flow.
1. Regulator or 
conserver is in “Off” 
position.
1. Set to prescribed 
setting.
2. Cylinder empty.
2. Replace/fill cylinder.
3. Toggle Valve closed.
3. Open toggle valve.
4. Conserver not 
sensing breath.
4a. Check position of 
cannula in nose. Do not 
breathe through mouth.
4b. Check to see if the 
cannula is kinked.