Invacare IRC5LXO2AW Manuale Utente

Pagina di 36
SECTION 7—TROUBLESHOOTING GUIDE
31
SECTION 7—TROUBLESHOOTING 
GUIDE
SYMPTOM:
PROBABLE CAUSE:
SOLUTION:
Alarm:
Main Power Loss:
Short beeps, long 
pause
1. Power cord NOT 
plugged in.
1. Insert plug into outlet.
Concentrator not 
operating,
power switch On.
2. No power at outlet.
2. Inspect house circuit 
breakers or fuses. If 
problem recurs, use a 
different outlet.
Beep......Beep......
3. Tripped circuit 
breaker.
3. Call Service Provider.
Alarm:
Continuous
Concentrator not 
operating, Power 
Switch On. 
Beeeeeeeep....
System Failure:
1. Unit overheating due 
to blocked air intake.
2. Insufficient power at 
outlet.
3. Internal repairs 
required
1a. Remove and clean 
cabinet filters.
1b. Move oxygen 
concentrator at least 7.5 cm 
(3 in) away from walls, 
draperies or furniture.
2. DO NOT use extension 
cords. Move to another 
electrical outlet or circuit.
3. Call Service Provider.