SYMPTOM:
PROBABLE CAUSE:
SOLUTION:
Alarm:
Main Power Loss:
Short beeps, long
pause
pause
1. Power cord NOT
plugged in.
plugged in.
1. Insert plug into outlet.
Concentrator not
operating,
power switch On.
operating,
power switch On.
2. No power at outlet.
2. Inspect house circuit
breakers or fuses. If
problem recurs, use a
different outlet.
breakers or fuses. If
problem recurs, use a
different outlet.
Beep......Beep......
3. Tripped circuit
breaker.
breaker.
3. Call Service Provider.
Alarm:
Continuous
Continuous
Concentrator not
operating, Power
Switch On.
operating, Power
Switch On.
Beeeeeeeep....
System Failure:
1. Unit overheating due
to blocked air intake.
1. Unit overheating due
to blocked air intake.
2. Insufficient power at
outlet.
outlet.
3. Internal repairs
required
required
1a. Remove and clean
cabinet filters.
cabinet filters.
1b. Move oxygen
concentrator at least 7.5 cm
(3 in) away from walls,
draperies or furniture.
concentrator at least 7.5 cm
(3 in) away from walls,
draperies or furniture.
2. DO NOT use extension
cords. Move to another
electrical outlet or circuit.
cords. Move to another
electrical outlet or circuit.
3. Call Service Provider.