SECTION 6—OPERATING INSTRUCTIONS
21
Perfecto
2
™Series
Flowrate
NOTE: For this procedure, refer to FIGURE 6.5.
NOTE: Take care not to set the flow above RED ring. An oxygen 
flow greater than 5 L/min will decrease the oxygen concentration.
1. Turn the flowrate knob to the setting prescribed by your 
physician or therapist.
 WARNING
DO NOT change the L/min setting on the flowmeter unless a 
change has been prescribed by your physician or therapist.
NOTE: To properly read the flowmeter, locate the prescribed 
flowrate line on the flowmeter. Next, turn the flow knob until the 
ball rises to the line. Now, center the ball on the L/min line 
prescribed. 
2. If the flowrate on the flowmeter ever falls below 0.5 L/min for 
more than about one minute, the LOW FLOW alarm will be 
triggered. This is a rapid beeping of the audible alarm. Check 
your tubing or accessories for blocked or kinked tubing or a 
defective humidifier bottle. After rated flow is restored to more 
than 0.5 L/min, the LOW FLOW audible alarm will go off.
NOTE: The use of some accessories such as the pediatric flowstand 
and the HomeFill compressor will deactivate the Low Flow Alarm.
FIGURE 6.5    Flowrate
Flow Knob
Ball
Flowmeter
Front Panel
0.5 L/min