FRx data storage
The FRx automatically stores data about its last clinical use in its 
internal memory. The stored data can be conveniently transferred to 
a personal computer or a handheld computer running the 
appropriate application in the Philips HeartStart Event Review data 
management software suite. Event Review software is for use by 
trained personnel only. Information about HeartStart Event Review 
is available online www.medical.philips.com/goto/eventreview.
Follow your local protocol with regard to prompt data transfer for 
medical review after using the FRx.
*
 Details about data transfer and 
timing are provided in Event Review documentation.
The information automatically stored by the FRx includes a summary 
of last-use data and detailed data about its last clinical use. You can 
get a voice summary of information about the last use of the 
defibrillator by holding the i-button down until it beeps once. The 
FRx will tell you how many shocks were delivered and how long it 
has been since it was turned on. Summary data are available anytime 
the defibrillator is ready for use (the battery and pads are installed, 
and the defibrillator is not turned on) or while it is actually in use. 
Removing the battery erases the summary data for the last use.
Last-use data stored in internal memory include:
ECG recordings (a maximum of 15 minutes following pads 
application
)
* The FRx automatically stores information about its last clinical use in its internal 
memory for at least 30 days, so the data can be downloaded to a computer 
running appropriate Event Review software. (If the battery is removed during 
this period, the defibrillator retains the files. When the battery is reinstalled, 
the last-use ECG recording will be kept in defibrillator memory for an additional 
30 days.) After this time, the last-use ECG recordings will automatically be 
erased to prepare for a future use.
† If ECG recordings from a previous use have not been erased, the maximum time 
for new ECG recordings may be less.