On/Off button
A green button located on the front of the HeartStart FRx Defibrillator. Pressing 
the On/Off button when the defibrillator is in standby mode turns the defibrillator 
on; pressing and holding the On/Off button for one second when the defibrillator 
is on turns the defibrillator off and disarms the defibrillator. In addition, pressing 
the On/Off button stops the battery insertion self-test that automatically runs 
when a battery is inserted.
pads
See “SMART Pads II.”
patient care pause
A defined period to allow CPR. See “NSA pause” and “protocol pause.”
periodic self-tests
Daily, weekly, and monthly tests automatically conducted by the HeartStart FRx 
Defibrillator when it is in its standby mode. The tests monitor many key functions 
and parameters of the defibrillator, including battery capacity, pads readiness, and 
the state of its internal circuitry.
protocol
A sequence of operations performed by the HeartStart FRx Defibrillator to direct 
patient care in the AED mode.
protocol pause
A period provided by the HeartStart FRx Defibrillator after a shock series, during 
which the responder can administer CPR. The defibrillator does not conduct 
background monitoring of the patient’s heart rhythm during this pause.
Quick Shock
The ability of the FRx to deliver a defibrillation shock very quickly – typically 
within 8 seconds – after the end of a patient care pause.
Ready light
A green LED showing the readiness for use of the HeartStart FRx Defibrillator. 
A blinking Ready light means the defibrillator is ready for use; a solid Ready light 
means the defibrillator is being used.
rhythm analysis
See “SMART analysis.”
Shock button
An orange button with a lightning bolt symbol on it, located on the front of the 
HeartStart FRx Defibrillator. The Shock button flashes when a shock is advised. 
You must press the button for the shock to be delivered.
shockable rhythm
A heart rhythm that the HeartStart FRx Defibrillator determines is appropriate 
for defibrillation, such as ventricular fibrillation and some ventricular tachycardias 
associated with sudden cardiac arrest. 
SMART analysis
The proprietary algorithm used by the HeartStart FRx Defibrillator to analyze the 
patient’s heart rhythm and determine whether a shock is advised.
SMART biphasic
waveform
The patented, low-energy defibrillation shock waveform used by the HeartStart 
FRx Defibrillator. It is an impedance-compensated biphasic waveform. It delivers 
150 Joules, nominal, into a 50 ohm load; when the Infant/Child Key is inserted, it 
delivers 50 Joules, nominal, into a 50 ohm load.